Pulse calibration
(Please refer also to the Bruker manual)
- Set D1 = 5, NS = 1, determine rg by rga,
record spectrum (zg) and correct phase
- Set P1 to 4x the old value (approx. 360 deg)
- ZG - EFP
- Adjust P1, ZG - EFP until approx. all
signals are zero (or as much positive as negative)
- The correct 90 deg pulse is the final value divided by 4.